The results of Hot Rod phantom with Xtrim PET system in PNP Co.

The results of Hot Rod phantom with Xtrim PET system in PNP Co.
Dedicated hot rod phantom was used for tomographic performance evaluation of the system

The main part of the phantom is a cylinder with 35 mm height and 32 mm diameter. Inside it, various fillable rods were drilled in 6 sections. The radius of rods ranges from 1.8 to 2.8 mm with steps of 2 mm from section to section. In each section, the distance between every 2 rods is equal to the radius of the rods in that section. The image of phantom without resolution recovery was optained

The results of Hot Rod phantom with Xtrim PET system in PNP Co.1

The results of Hot Rod phantom with Xtrim PET system in PNP Co.2

The results of Image Quality phantom with Xtrim PET system in PNP Co.

The results of Hot Rod phantom with Xtrim PET system in PNP Co.3

The results of Hot Rod phantom with Xtrim PET system in PNP Co.4